Ozone Output Chart For The New AOS-1M With Fan
LPM | cc/min | 5% | 10% | 20% | 30% | 40% | 50% | 60% | 70% | 80% | 90% | 100% |
1/32 1/16 1/8 1/4 1/2 |
31 62 125 250 500 |
79 60 46 25 13 |
80 61 48 26 14 |
84 64 50 27 15 |
87 66 52 29 16 |
90 70 55 31 17 |
94 74 58 34 19 |
97 78 62 37 21 |
100 83 67 41 23 |
105 194 75 48 27 |
110 103 86 60 37 |
111 109 95 71 45 |
Please Note! Outputs from the prototype, when more are tested we will post an average.
Note! Output is in gamma.
AOS-1MS with fan outputs tested with Puritan Bennett low flow regulator.
O2 liters/min |
VC 1 |
VC 1.5 |
VC 2 |
VC 2.5 |
VC 3 |
VC 3.5 |
VC 4 |
1/32 |
99 |
105 |
112 |
109 |
110 |
101 |
90 |
Important Notes: When medical ozone concentration factors are given by various manufactures the following will reduce your confusion: 1 ug/ml = 1 mg/l = 1 g/m3 = 1 gamma. Also, 1 ml is = to 1 cc.
For instance, the output for the AOS-1MS at 1/2 liter/min flow with the number 4 setting of the variable control (VC), the output equals 68ug/cc = 68ug/ml = 68mg/l = 68g/m3 = 68 gamma.
Factors that can influence the outputs are elevation, humidity, pressure, temperature or the quality of oxygen being used. There can also be a variance of outputs between different brands of oxygen regulators and flow meters.
Note: 1 mcg/ml (microgram per milliliter) is equal to 1 ug/ml.